External Quality Assurance (EQA) Program

for

G6PD Quantitative Test in Taiwan


Survey No. : RH2013-07 Sample sent on : 2013/07/01 Reporting deadline : 2013/07/08
This Survey S2 has been used in 2009-2013 (n = 6; Survey / Sample No.) :
RH2010-03/S3 ; RH2012-11/S1 ; RH2013-07/S2

Intermediate Measurement Precision of G6PD Quantitative Test

Lab

Survey : RH2013-07 / Sample 2

Period : 2010 ~ 20134

U/gHb

SDI2

Mean

n

SD

CV

RH01

9.2
-0.1
9.3
3
1.0
10.8%

RH02

9.8
-0.5
10.1
3
0.6
5.9%

RH03

10.3
-0.2
10.4
3
0.6
5.8%

RH04

9.6
1.0
9.5
3
0.1
1.1%

RH06

9.6
0.0
9.6
3
0.3
3.1%

RH07

9.9
0.3
9.8
3
0.3
3.1%

RH08

9.3
-0.4
9.5
3
0.5
5.3%

RH09

10.4
1.0
9.8
3
0.6
6.1%

RH10

9.7
-0.7
10.1
3
0.6
5.9%

RH12

9.6
-1.0
9.9
3
0.3
3.0%

RH13

10.0
0.0
10
3
0.6
6.0%

RH14

9.8
0.5
9.6
3
0.4
4.2%

RH18

10.4
0.7
10
3
0.6
6.0%

RH19

9.8
-0.8
10.3
3
0.6
5.8%

A0189

11.8
1.1
10.7
3
1.0
9.3%

A0203

9.3
-1.3
9.8
3
0.4
4.1%

G026

10.4
1.0
9.9
3
0.5
5.1%

CL001

9.2
0.0
9.2
3
0.4
4.3%

CL002

9.4
-0.4
9.8
3
0.9
9.2%

B001

9.2
-0.2
9.5
3
1.9
20.0%

CL014

11.3
-
11.3
1
-
-

CL015B

9.6
-
9.6
1
-
-

Total participating laboratories = 22

Median

9.8
-0.1
9.8
-
0.6
5.8%

Range

9.2 ~ 11.8
-1.3 ~ 1.1
9.2 ~ 11.3
-
0.1 ~ 1.9
1.1% ~ 20.0%

 

1).
: Outlier (this result has been dropped and was not included in calculation of Mean, SD and CV).
2). Standard Deviation Index (SDI) = (Individual Result - Individual Mean) / Individual SD
3). The reference values were determined by reference method at the QC reference Lab.
4). Survey / Sample No.¡GRH2010-03/S3 ; RH2012-11/S1 ; RH2013-07/S2
*Intermediate Measurement Precision¡FISO 3534-2:2006

 


Contact us
Address¡G P.O. Box 26-624, Taipei, Taiwan, 10699¡@¡@Tel¡G+886-2-2703-6080
  ¡@¡@e-mail¡G
¡@Fax¡G +886-2-2703-6070

     
                  Preventive Medicine Foundation, Taipei, Taiwan

updated : 30 Jul, 2013