Survey No. : RH2018-01 | Sample sent on : 2018 / 02 / 26 | Reporting deadline : 2018 / 03 / 05 |
This
Survey S2
has been used in 2013-2018 (n = 6; Survey / Sample No.) : RH2013-05/S1 ; RH2015-07/S1 ; RH2016-07/S2 ; RH2016-10/S3 ; RH2017-01/S2 ; RH2018-01/S3 |
Intermediate Measurement Precision of G6PD Quantitative Test
Lab |
Survey : RH2018-01 / Sample 3 |
Period : 2013 ~ 20182 |
|||||
U/gHb |
SDI1 |
Mean |
n |
SD |
CV |
||
RH01 |
3.7 |
0.0 |
3.7 |
6 |
0.45 |
12.2% |
|
RH02 |
3.7 |
-0.9 |
4.1 |
6 |
0.46 |
11.2% |
|
RH04 |
3.8 |
0.0 |
3.8 |
6 |
0.17 |
4.5% |
|
RH06 |
4.1 |
0.7 |
3.8 |
6 |
0.44 |
11.6% |
|
RH07 |
3.4 |
-1.1 |
3.8 |
6 |
0.38 |
10.0% |
|
RH08 |
3.7 |
0.4 |
3.6 |
6 |
0.27 |
7.5% |
|
RH09 |
3.3 |
-1.5 |
3.7 |
6 |
0.26 |
7.0% |
|
RH10 |
4.0 |
0.7 |
3.8 |
6 |
0.27 |
7.1% |
|
RH12 |
3.2 |
-0.4 |
3.4 |
6 |
0.48 |
14.1% |
|
RH13 |
3.8 |
-0.3 |
3.9 |
6 |
0.40 |
10.3% |
|
RH14 |
3.7 |
0.0 |
3.7 |
6 |
0.49 |
13.2% |
|
RH19 |
3.0 |
-1.1 |
3.4 |
6 |
0.35 |
10.3% |
|
A0203 |
3.7 |
0.2 |
3.6 |
6 |
0.46 |
12.8% |
|
G026 |
3.6 |
-0.3 |
3.7 |
6 |
0.35 |
9.5% |
|
CL001 |
3.3 |
-0.8 |
3.8 |
6 |
0.60 |
15.8% |
|
CL002 |
3.9 |
0.4 |
3.8 |
6 |
0.28 |
7.4% |
|
CL014 |
3.4 |
-0.4 |
3.5 |
6 |
0.28 |
8.0% |
|
CL015B |
3.6 |
0.3 |
3.5 |
6 |
0.40 |
11.4% |
|
Total participating laboratories = 21 |
|||||||
Median |
3.9 |
0.4 |
3.7 |
¢w |
0.4 |
11.1% |
|
Range |
3.2 ~ 4.4 |
-0.97 ~ 1.54 |
3.4 ~ 4.1 |
¢w |
0.17 ~ 0.79 |
4.6% ~ 19.3% |
2).
Survey / Sample No.¡GRH2013-05/S1 ; RH2015-07/S1 ; RH2016-07/S2 ; RH2016-10/S3 ; RH2017-01/S2 ; RH2018-01/S3 |
*Intermediate Measurement Precision¡FISO 3534-2:2006, JCGM 200:2012 ( VIM, 3rd ed. ) 2.23 |
** Mean and SD are calculated by Robust method according to ISO 13528 |
*** n < 5, not calculated Mean and SD.
|
Contact us
Address¡G
P.O. Box 26-624, Taipei, Taiwan, 10699¡@¡@Tel¡G+886-2-2703-6080
¡@¡@e-mail¡G
¡@Fax¡G
+886-2-2703-6070
Preventive
Medicine Foundation, Taipei, Taiwan
updated : 13 Mar, 2018