External Quality Assurance (EQA) Program

for

G6PD Quantitative Test in Taiwan


Survey No. : RH2015-01 Sample sent on : 2015/01/26 Reporting deadline : 2015/02/02
This Survey S1 has been used in 2009-2015 (n = 8; Survey / Sample No.) :
RH2009-11/S1 ; RH2010-01/S2 ; RH2011-03/S2 ; RH2011-07/S1 ; RH2012-09/S2 ; RH2013-05/S2 ; RH2014-04/S2 ; RH2015-01/S1

Intermediate Measurement Precision of G6PD Quantitative Test

Lab

Survey : RH2015-01 / Sample 1

Period : 2009 ~ 20154

U/gHb

SDI2

Mean

n

SD

CV

RH01

8.7
0.8
7.8
8
1.1
14.1%

RH02

8.4
0.2
8.2
8
0.7
9.1%

RH04

8.2
-0.6
8.4
8
0.3
4.1%

RH06

7.7
0.0
8.3
8
0.5
5.7%

RH07

7.9
-0.7
8.3
8
0.6
6.7%

RH08

8.0
-0.6
8.3
8
0.5
5.8%

RH09

8.8
1.1
8.3
8
0.5
6.0%

RH10

7.7
-1.3
8.2
8
0.4
4.4%

RH12

7.8
-0.8
8.3
8
0.6
7.0%

RH13

8.6
1.0
8.3
8
0.3
3.2%

RH14

7.5
-1.5
8.2
8
0.4
5.5%

RH18

8.9
0.7
8.4
8
0.8
9.3%

RH19

7.9
-0.9
8.6
8
0.8
9.2%

A0189

8.2
-0.1
8.3
8
0.9
11.4%

A0203

8.0
-0.6
8.3
8
0.6
6.8%

G026

8.4
-0.4
8.6
8
0.5
5.5%

CL001

8.6
0.8
8.2
8
0.6
7.2%

CL002

7.8
-1.0
8.3
8
0.4
5.2%

B001

7.8
-0.5
8.1
8
0.7
8.3%

CL014

8.2
-0.6
8.4
6
0.3
4.1%

CL015B

8.5
0.7
8.1
3
0.5
6.6%

Total participating laboratories = 21

Median

8.2
-0.5
8.3
-
0.5
6.6%

Range

7.5 ~ 8.9
-1.5 ~ 1.1
7.8 ~ 8.6
-
0.3 ~ 1.1
3.2% ~ 14.1%

 

1). Standard Deviation Index (SDI) = (Individual Result - Individual Mean) / Individual SD
*Intermediate Measurement Precision¡FISO 3534-2:2006
** Mean and SD are calculated by Robust method according to ISO 13528:2005

 


Contact us
Address¡G P.O. Box 26-624, Taipei, Taiwan, 10699¡@¡@Tel¡G+886-2-2703-6080
  ¡@¡@e-mail¡G
¡@Fax¡G +886-2-2703-6070

     
                  Preventive Medicine Foundation, Taipei, Taiwan

updated : 06 Feb, 2015


[ Home ]