External Quality Assurance (EQA) Program

for

G6PD Quantitative Test in Philippines


Survey No. : RH2010-03 Sample sent on : 2010/03/09 Reporting deadline : 2010/03/16
This Survey S2 has been used in 2009-2010 (n = 2 ; Survey / Sample No.) : RH 2009-07/S2 ; RH 2010-03/S2 ;

Intermediate Measurement Precision of G6PD Quantitative Test

Lab

Survey : RH2010-03 / Sample 2

Program period: 2009 ~ 20104

U/gHb

SDI2

Mean

n

SD

CV

F01

4.9

4.9

15

F02

5.0

5.2

2

F03

3.9

4.8

2

F04

4.9

5.3

2

F05

6

5.9

16

F06

6.4

6.3

2

F07

3.9

4.8

2

F08

3.8

4.9

2

F09

6

3.4

16

F10

4.3

5.4

2

Total participating laboratories = 10

Median

4.6

5.1

Range

3.8 ~ 6.4

3.4 ~ 6.3

Ref. Lab3

5.1

0.0

5.1

2

0.27

5.3%

1).
 
: Outlier (this result has been dropped and was not included in calculation of Mean, SD and CV).
2). Standard Deviation Index (SDI) = (Individual Result - Individual Mean) / Individual SD
3). The reference values were determined by reference method at the QC reference Lab.
4). Survey / Sample No.RH 2009-07/S2RH 2010-03/S2
5). Not participated in the RH2009-07 survey.
6). Results not reported for the RH2010-03 survey.
*. Intermediate Measurement Precision;ISO 3534-2:2006

[ Home ]

Newborn Screening QA Program

Preventive Medicine Foundation

Newborn Screening Reference Center

National Institutes of Health

P.O. Box 624, Taipei Xinwei, Taiwan 10699 University of the Philippines Manila  
Tel : + 886-2-2703-6080;  Fax : +886-2-2703-6070  

Unit 104, Building A, UP Ayala Land Technohub, Brgy. UP Campus,
Diliman, Quezon City 1101, Philippines

 
  Fax : + 632-351-7903;  
         