Survey No. : RH2018-02 | Sample sent on : 2018 / 04 / 23 | Reporting deadline : 2018 / 04 / 30 |
This
Survey S1
has been used in 2013-2018 (n = 5; Survey / Sample No.) : RH2013-01/S1 ; RH2013-07/S3 ; RH2014-04/S1 ; RH2017-01/S1 ; RH2018-02/S1 |
Intermediate Measurement Precision of G6PD Quantitative Test
Lab |
Survey : RH2018-02 / Sample 1 |
Period : 2013 ~ 20182 |
|||||
U/gHb |
SDI1 |
Mean |
n |
SD |
CV |
||
RH01 |
21.9 |
2.2 |
17.6 |
5 |
1.98 |
11.3% |
|
RH02 |
14.6 |
-1.5 |
16.1 |
5 |
1.00 |
6.2% |
|
RH04 |
18.5 |
0.2 |
18.2 |
5 |
1.26 |
6.9% |
|
RH06 |
19.0 |
9.0 |
17.2 |
5 |
0.20 |
1.2% |
|
RH07 |
15.6 |
-1.4 |
16.6 |
5 |
0.72 |
4.3% |
|
RH08 |
17.8 |
0.7 |
17.2 |
5 |
0.84 |
4.9% |
|
RH09 |
18.3 |
0.9 |
17.6 |
5 |
0.81 |
4.6% |
|
RH10 |
17.8 |
1.1 |
17.1 |
5 |
0.63 |
3.7% |
|
RH12 |
17.3 |
-0.7 |
18.1 |
5 |
1.13 |
6.2% |
|
RH13 |
19.1 |
1.6 |
17.9 |
5 |
0.75 |
4.2% |
|
RH14 |
17.8 |
0.9 |
17.1 |
5 |
0.80 |
4.7% |
|
RH19 |
16.2 |
-0.6 |
17.1 |
5 |
1.43 |
8.4% |
|
A0203 |
16.1 |
-0.8 |
17.0 |
5 |
1.06 |
6.2% |
|
G026 |
16.5 |
-0.5 |
17.3 |
5 |
1.54 |
8.9% |
|
CL001 |
17.5 |
0.8 |
16.9 |
5 |
0.75 |
4.4% |
|
CL002 |
18.7 |
0.3 |
18.5 |
5 |
0.79 |
4.3% |
|
CL014 |
19.3 |
0.7 |
18.4 |
5 |
1.31 |
7.1% |
|
CL015B |
16.3 |
¢w |
¢w |
4 |
¢w |
¢w |
|
Total participating laboratories = 18 |
|||||||
Median |
17.8 |
0.7 |
17.2 |
¢w |
0.84 |
4.9% |
|
Range |
14.6 ~ 21.9 |
-1.5 ~ 9.0 |
16.1 ~ 18.5 |
¢w |
0.20 ~ 1.98 |
1.2% ~ 11.3% |
*Intermediate Measurement Precision¡FISO 3534-2:2006, JCGM 200:2012 ( VIM, 3rd ed. ) 2.23 |
** Mean and SD are calculated by Robust method according to ISO 13528 |
*** n < 5, not calculated Mean and SD.
|
Contact us
Address¡G
P.O. Box 26-624, Taipei, Taiwan, 10699¡@¡@Tel¡G+886-2-2703-6080
¡@¡@e-mail¡G
¡@Fax¡G
+886-2-2703-6070
Preventive
Medicine Foundation, Taipei, Taiwan
updated : 08 May, 2018