Intermediate Measurement Precision of G6PD Quantitative Test

Lab

Survey : RH2013-05 / Sample 2

Period : 2009 ~ 20134

U/gHb

SDI2

Mean

n

SD

CV

RH01

6.9
-0.8
7.8
6
1.1
14.1%

RH02

10.4
1.8
8.4
6
1.1
13.1%

RH03

8.3
0.0
8.3
6
0.8
9.6%

RH04

8.2
-0.7
8.4
6
0.3
3.6%

RH06

8.1
-0.8
8.4
6
0.4
4.8%

RH07

8.0
-0.5
8.2
6
0.4
4.9%

RH08

8.0
-0.6
8.3
6
0.5
6.0%

RH09

8.8
1.3
8.3
6
0.4
4.8%

RH10

8.0
-1.0
8.4
6
0.4
4.8%

RH12

8.0
-0.5
8.4
6
0.8
9.5%

RH13

8.3
0.0
8.3
6
0.2
2.4%

RH14

8.3
0.3
8.2
6
0.3
3.7%

RH18

9.7
1.6
8.4
6
0.8
9.5%

RH19

8.4
-0.6
8.8
6
0.7
8.0%

A0189

7.8
-0.5
8.5
6
1.3
15.3%

A0203

8.4
0.2
8.2
6
0.9
11.0%

G026

9.3
1.5
8.7
6
0.4
4.6%

CL001

8.5
0.7
8.1
6
0.6
7.4%

CL002

8.0
-1.3
8.4
6
0.3
3.6%

B001

8.6
0.3
8.4
6
0.6
7.1%

CL014

9.3
1.4
8.6
4
0.5
5.8%

CL015B

7.6
-
7.6
1
-
-

Total participating laboratories = 22

Median

8.3
0.0
8.4
0.5
6.0%

Range

6.9 ~ 10.4
-1.3 ~ 1.8
7.6 ~ 8.8
0.2 ~ 1.3
2.4% ~ 15.3%

 

1).
: Outlier (this result has been dropped and was not included in calculation of Mean, SD and CV).
2). Standard Deviation Index (SDI) = (Individual Result - Individual Mean) / Individual SD
3). The reference values were determined by reference method at the QC reference Lab.
*Intermediate Measurement Precision¡FISO 3534-2:2006

 


Contact us