Survey No. : RH2011-01 | Sample sent on : 2011/01/17 | Reporting deadline : 2011/01/24 |
This
Survey S3
has been used in 2007-2011 (n = 8 ; Survey / Sample No.) :
RH 2007-03/S2 ; RH 2007-07/S1 ; RH 2007-09/S1 ; RH 2008-03/S2 ; RH 2008-11/S2 ; RH 2009-07/S2 ; RH 2010-03/S2 ; RH 2011-01/S3 |
Intermediate Measurement Precision of G6PD Quantitative Test
Lab |
Survey : RH2011-01 / Sample 2 |
Period : 2007 ~ 20114 |
|||||
U/gHb |
SDI2 |
Mean |
n |
SD |
CV |
||
RH01 |
6.2 |
1.5
|
5.5 |
8 |
0.46 |
8.4% |
|
RH02 |
4.8 |
-0.8
|
5.1 |
8 |
0.39 |
7.6% |
|
RH03 |
5.6 |
0.8
|
5.1 |
8 |
0.66 |
12.9% |
|
RH04 |
5.7 |
1.1
|
5.0 |
8 |
0.66 |
13.2% |
|
RH06 |
5.2 |
0.3
|
5.1 |
8 |
0.38 |
7.5% |
|
RH07 |
5.1 |
0.5
|
4.7 |
8 |
0.80 |
17.0% |
|
RH08 |
5.2 |
-0.3
|
5.3 |
8 |
0.33 |
6.2% |
|
RH09 |
5.3 |
0.9
|
4.9 |
8 |
0.43 |
8.8% |
|
RH10 |
5.4 |
-0.2
|
5.5 |
8 |
0.48 |
8.7% |
|
RH12 |
4.5 |
-0.8
|
5.2 |
7* |
0.54 |
10.4% |
|
RH13 |
5.2 |
0.4
|
5.0 |
8 |
0.51 |
10.2% |
|
RH14 |
5.4 |
0.6
|
5.1 |
8 |
0.48 |
9.4% |
|
RH18 |
5.2 |
0.5
|
4.9 |
8 |
0.60 |
12.2% |
|
RH19 |
4.8 |
-1.0
|
5.1 |
8 |
0.29 |
5.4% |
|
A0189 |
5.1 |
0.4
|
4.8 |
8 |
0.83 |
17.3% |
|
A0203 |
5.7 |
1.1
|
5.1 |
8 |
0.53 |
10.4% |
|
G031 |
6.4 |
1.6
|
5.4 |
8 |
0.64 |
11.9% |
|
G026 |
4.9 |
-1.3
|
5.3 |
8 |
0.32 |
6.0% |
|
CL01 |
4.7 |
-0.8
|
4.9 |
8 |
0.26 |
5.3% |
|
CL02 |
4.7 |
-0.2
|
4.8 |
8 |
0.55 |
11.5% |
|
B001 |
4.6 |
-1.1
|
5.3 |
8 |
0.66 |
12.5% |
|
Total participating laboratories = 21 |
|||||||
Median |
5.1 |
0.4 |
5.1 |
¡Ð |
0.77 |
9.8% |
|
Range |
4.5 ~ 6.4 |
-1.3 ~ 1.6 |
4.7 ~ 5.8 |
¡Ð |
0.26 ~ 0.83 |
5.3% ~ 17.3% |
|
Ref. Lab3 |
5.1 |
0.0 |
5.1 |
8 |
0.25 |
4.9% |
1).
|
: Outlier (this result has been dropped and was
not included in calculation of Mean, SD and CV).
|
|||
3).
The reference values were determined by reference method at the
QC reference Lab.
|
||||
4).
Survey / Sample No.¡GRH
2007-03/S2 ; RH
2007-07/S1 ; RH
2007-09/S1 ; RH
2008-03/S2 ; RH
2008-11/S2 ; RH
2009-07/S2 ; RH
2010-03/S2 ; RH
2011-01/S3
|
||||
*
An error result of RH2009-07 was excluded. |
||||
**Intermediate Measurement Precision¡FISO 3534-2:2006
|
Contact us
Address¡G
P.O. Box 26-624, Taipei, Taiwan, 10699¡@¡@Tel¡G+886-2-2703-6080
¡@¡@e-mail¡G
¡@Fax¡G
+886-2-2703-6070
Preventive
Medicine Foundation, Taipei, Taiwan
updated : 30 Jul, 2013