Survey No. : RH2010-05 | Sample sent on : 2010/05/03 | Reporting deadline : 2010/05/10 |
This
Survey S3
has been used in 2007-2010 (n = 7 ; Survey / Sample No.) :
RH 2007-01/S2 ; RH 2007-07/S2 ; RH 2007-11/S3 ; RH 2008-01/S3 ; RH 2008-07/S2 ; RH 2009-01/S1 ; RH 2010-05/S3 |
Intermediate Measurement Precision of G6PD Quantitative Test
Lab |
Survey : RH2010-05 / Sample 3 |
Period : 2007 ~ 20104 |
||||||
U/gHb |
SDI2 |
Mean |
n |
SD |
CV |
|||
RH01 |
13.8 |
0.4 |
13.5 |
7 |
0.80 |
5.9% |
||
RH02 |
13.0 |
0.6 |
12.8 |
7 |
0.35 |
2.7% |
||
RH03 |
13.4 |
0.5 |
12.9 |
7 |
0.92 |
7.1% |
||
RH04 |
12.0 |
-0.8 |
12.6 |
7 |
0.73 |
5.8% |
||
RH06 |
12.5 |
-0.8 |
13.1 |
7 |
0.73 |
5.6% |
||
RH07 |
12.1 |
-0.9 |
12.9 |
7 |
0.85 |
6.6% |
||
RH08 |
12.5 |
-1.6 |
13.5 |
7 |
0.61 |
4.5% |
||
RH09 |
11.6 |
-1.2 |
12.5 |
7 |
0.73 |
5.8% |
||
RH10 |
13.4 |
0.0 |
13.4 |
7 |
0.72 |
5.4% |
||
RH12 |
12.9 |
0.2 |
12.6 |
7 |
1.26 |
10.0% |
||
RH13 |
11.7 |
-1.4 |
13.4 |
7 |
1.21 |
9.0% |
||
RH14 |
12.5 |
-0.3 |
13.0 |
7 |
1.47 |
11.3% |
||
RH18 |
13.4 |
1.2 |
12.5 |
7 |
0.78 |
6.2% |
||
RH19 |
13.7 |
0.8 |
13.1 |
7 |
0.76 |
5.8% |
||
A0189 |
13.9 |
0.8 |
12.9 |
7 |
1.25 |
9.7% |
||
A0203 |
12.4 |
-0.3 |
12.7 |
7 |
0.98 |
7.7% |
||
G031 |
12.8 |
-0.8 |
13.6 |
7 |
1.03 |
7.6% |
||
G026 |
12.7 |
-0.7 |
13.7 |
7 |
1.35 |
9.9% |
||
CL01 |
11.7 |
-1.2 |
12.4 |
7 |
0.60 |
4.8% |
||
CL02 |
13.5 |
0.5 |
13.1 |
7 |
0.87 |
6.6% |
||
B001 |
13.5 |
0.8 |
12.9 |
7 |
0.78 |
6.0% |
||
Total participating laboratories = 21 |
||||||||
Median |
12.8 |
-0.3 |
12.9 |
- |
0.80 |
6.2% |
||
Range |
11.6 ~ 13.9 |
-1.6 ~ 1.2 |
12.4 ~ 13.7 |
- |
0.35 ~ 1.47 |
2.7% ~ 11.3% |
||
Ref. Lab3 |
12.7 |
0.0 |
12.7 |
7 |
0.25 |
2.0% |
1).
|
: Outlier
(this result has been dropped and was not included in calculation
of Mean, SD and CV).
|
|||
3).
The reference values were determined by reference method at the
QC reference Lab.
|
||||
4).
Survey / Sample No.:RH
2007-01/S2 ; RH
2007-07/S2 ; RH
2007-11/S3 ; RH
2008-01/S3 ; RH
2008-07/S2 ; RH
2009-01/S1 ; RH
2010-05/S3 |
||||
*Intermediate Measurement Precision;ISO 3534-2:2006
|
Contact us
Address:
P.O. Box 26-624, Taipei, Taiwan, 10699 Tel:+886-2-2703-6080
e-mail:
Fax:
+886-2-2703-6070
Preventive
Medicine Foundation, Taipei, Taiwan
updated : 30 Jul, 2013