Survey No. : RH2013-07 | Sample sent on : 2013 / 07 / 01 | Reporting deadline : 2013 / 07 / 08 |
This
Survey S2
has been used in 2010-2013 (n = 3 ; Survey / Sample No.) :
RH2010-03/S3 ; RH2012-11/S1 ; RH2013-07/S2 ; |
Intermediate Measurement Precision of G6PD Quantitative Test
Lab |
Survey : RH2013-07 / Sample 2 |
Program period: 2010 ~ 20133 |
|||||
U/gHb |
SDI2 |
Mean |
n |
SD |
CV |
||
F01 |
5.8 |
-1.1 |
7.9 |
3 |
1.9 |
24.1% |
|
F02 |
5.3 |
-1.0 |
7.8 |
3 |
2.6 |
33.3% |
|
F03 |
5.7 |
-1.0 |
7.1 |
3 |
1.4 |
19.7% |
|
F04 |
4.9 |
-1.2 |
6.9 |
3 |
1.7 |
24.6% |
|
F05 |
4.5 |
-0.7 |
5.9 |
2 |
2.0 |
33.9% |
|
F06 |
4.8 |
-0.7 |
5.8 |
2 |
1.4 |
24.1% |
|
F07 |
7.7 |
-1.3 |
8.1 |
3 |
0.3 |
3.7% |
|
F08 |
4.7 |
-1.1 |
6.6 |
3 |
1.7 |
25.8% |
|
F09 |
5.5 |
-0.7 |
6.3 |
2 |
1.1 |
17.5% |
|
F10 |
4.0 |
-1.0 |
6.3 |
3 |
2.3 |
36.5% |
|
F11 |
3.7 |
- |
5.3 |
2 |
2.3 |
43.4% |
|
F12 |
4.4 |
- |
5.6 |
2 |
1.7 |
30.4% |
|
F13 |
5.7 |
- |
5.7 |
1 |
- |
- |
|
F14 |
3.6 |
- |
3.6 |
1 |
- |
- |
|
F15 |
4.1 |
- |
4.1 |
1 |
- |
- |
|
Total participating laboratories = 10 |
|||||||
Median |
4.8 |
-1.0 |
6.3 |
- |
1.7 |
25.2% |
|
Range |
3.6 ~ 7.7 |
-1.3 ~ -0.7 |
3.6 ~ 8.1 |
- |
0.3 ~ 2.6 |
3.7% ~ 43.4% |
1).
|
: Outlier
(this result has been dropped and was not included in calculation
of Mean, SD and CV).
|
|
*.
Intermediate Measurement Precision;ISO 3534-2:2006
|
Preventive
Medicine Foundation
Unit 104, Building A, UP Ayala Land Technohub, Brgy. UP Campus,
Newborn
Screening QA Program
P.O. Box 624, Taipei Xinwei, Taiwan 10699
University
of the Philippines Manila
Tel
: + 886-2-2703-6080;
Fax
: +886-2-2703-6070
Diliman, Quezon City 1101, Philippines
Fax
: + 632-351-7903;
updated