Survey No. : RH2011-05 Sample sent on : 2011 年 05 月 09 日 Reporting deadline : 2011 月 05 月 16 日
This Survey S2 has been used in 2006-2010 (n = 6 ; Survey / Sample No.) :
RH 2006-09/S1 ; RH 2006-10/S1 ; RH 2007-03/S3 ; RH 2008-05/S2 ; RH 2010-01/S1 ; RH 2011-05/S2

Intermediate Measurement Precision of G6PD Quantitative Test

Lab

Survey : RH2011-03 / Sample 1

Period : 2006 ~ 20114

U/gHb

SDI2

Mean

n

SD

CV

RH01

22.1
0.3
21.4
6
2.55
11.9%

RH02

20.0
-0.6
20.6
6
1.06
5.1%

RH03

20.6
0.5
19.6
6
1.84
9.4%

RH04

18.4
-0.7
19.4
6
1.47
7.6%

RH06

21.6
1.6
20.3
6
0.81
4.0%

RH07

19.7
0.3
19.4
6
1.04
5.4%

RH08

22.0
1.9
20.4
6
0.86
4.2%

RH09

18.9
-1.3
20.5
6
1.23
6.0%

RH10

19.0
-1.5
21.6
6
1.77
8.2%

RH12

21.6
0.2
21.4
6
1.12
5.2%

RH13

18.9
-1.3
20.3
6
1.10
5.4%

RH14

21.5
0.5
20.7
6
1.68
8.1%

RH18

20.5
0.6
19.8
6
1.20
6.1%

RH19

19.0
-1.1
19.8
6
0.75
3.8%

A0189

22.4
0.6
21.5
5*
1.40
6.5%

A0203

19.7
-0.8
20.1
6
0.48
2.4%

G026

20.1
-0.1
20.3
6
1.48
7.3%

CL01

19.3
-0.5
19.6
6
0.55
2.8%

CL02

20.9
-0.1
21.0
6
1.08
5.1%

B001

21.8
0.9
20.3
6
1.60
7.9%

Total participating laboratories = 20

Median

20.3

0.0

20.3

1.20

5.7%

Range

18.4 ~ 22.4

-1.5 ~ 1.9

19.4 ~ 21.6

0.48 ~ 2.55

2.4% ~ 11.9%

Ref. Lab3

20.5

0.6

20.3

6

0.36

1.8%

1).
: Outlier (this result has been dropped and was not included in calculation of Mean, SD and CV).
2). Standard Deviation Index (SDI) = (Individual Result - Individual Mean) / Individual SD
3). The reference values were determined by reference method at the QC reference Lab.
* An error result of RH2010-01 was excluded.
**Intermediate Measurement Precision;ISO 3534-2:2006