Sample sent on : 2006-09-11 | Report deadline : 2006-09-18 | |
Total participating laboratories : 14 | Total reports returned : 14 |
Lab. | Receive (day) |
Test (day) |
Report (day) |
Sample 1 |
Sample 2 |
Sample 3 |
Sample 4 |
Sample 5 |
Sample 6 |
Sample 7 |
Sample 8 |
Sample 9 |
Sample 10 |
Cut-Off Value |
Reagent Code |
A01 | 0 | 2 | 2 | >7.8 | >7.8 | 2.0(P) | 3.5 | 3.4 | >7.8 | 3.4 | 2.1(P) | 1.8(P) | >7.8 | 2.9 | 1 |
A02 | 0 | 0 | 1 | 7.5 | 7.4 | 1.5(P) | 3.2 | 3.3 | 7.4 | 3.4 | 1.6(P) | 1.6(P) | 7.5 | 2.6 | 1 |
A03 | 0 | 0 | 1 | 23.5 | 24.5 | 3.5(P) | 8.1 | 8.1 | 24.4 | 9.2 | 3.9(P) | 3.9(P) | 24.3 | 6.0 | 6 |
B01 | 2 | 3 | 3 | + | + | +/-(P) | + | + | + | + | +/-(P) | +/-(P) | + | none | 7 |
B02 | 4 | 7 | 7 | - | - | +(P) | - | - | - | - | +(P) | +(P) | - | none | 99 |
B03 | 2 | 2 | 15 | 3.83 | 3.89 | 1.22(P) | 2.51 | 2.31 | 4.51 | 2.36 | 1.15(P) | 1.25(P) | 4.06 | 2.2 | 5 |
B04 | 9 | 9 | 10 | 7.0 | 6.8 | 1.1(P) | 2.5 | 2.6 | 6.8 | 2.3 | 1.0(P) | 1.1(P) | 7.0 | 2.0 | 1 |
C01 | 2 | 3 | 4 | 7.6 | 7.4 | 1.4(P) | 2.7 | 2.6 | 7.2 | 2.8 | 1.4(P) | 1.2(P) | 7.2 | 1.8 | 1 |
C02 | 2 | 7 | 7 | - | - | +(P) | - | - | - | - | +(P) | +(P) | - | none | 4 |
C04 | 4 | 4 | 10 | 16.9 | 15.4 | 2.0(P) | 7.2 | 7.0 | 16.4 | 7.2 | 2.1(P) | 2.9(P) | 16.0 | 4.4 | 5 |
C05 | 2 | 3 | 7 | 19,03 | 17,63 | 5,68 | 8,43 | 8,15 | 18,95 | 8,50 | 3,81 | 5,0 | 17,51 | 3,55 | 3 |
D01 | 3 | 7 | 8 | 6.75 | 6.85 | 0.95(P) | 1.15(P) | 1.50(P) | 7.10 | 1.05(P) | 1.25(P) | 1.10(P) | 7.25 | 2.6 | 1 |
D02 | 4 | 4 | 4 | 6.17 | 6.39 | 1.02(P) | 1.51(P) | 1.59(P) | 5.71 | 1.65(P) | 0.77(P) | 0.9(P) | 6.1 | 2.19 | 2 |
D03 | 2 | 4 | 4 | 15,6 | 15,0 | 1,9(P) | 2,5(P) | 2,4(P) | 17,3 | 2,8(P) | 1,3(P) | 1,7(P) | 16,5 | 4 | 11 |
Refer- ence* |
- | - | - | 28.2 | 28.2 | 3.0 (P) | 7.0 | 7.0 | 28.2 | 7.0 | 3.0 (P) | 3.0 (P) | 28.2 | 4.4 |   |
Median (Range) |
2 (0-9) |
3.5 (0-9) |
5.5 (1-15) |
* The reference values were determined by using Trinity Biotech 345-UV reagent at the QC reference Lab.
** The positive result assigned by the participants are indicated by red color in the table and the (P) indicates out-of-expected range (i.e. Positive) decision.
Reagent Code |
Laboratory |
|
Ani Labsystems Neonatal G6PD | 11 | D03 |
Bio-Rad | 2 | D02 |
Gamma (G6PD-2000) | 5 | B03, C04 |
Micky | 7 | B01 |
PE (ND-1000EY) | 1 | A01, A02, B04, C01, D01 |
R&D DIAGNOSTICS (OSMMR2000-D) | 3 | C05 |
R&D DIAGNOSTICS (OSMMR720) | 4 | C02 |
Trinity Biotech 345UV | 6 | CBRL, A03 |
Laboratory prepared | 99 | B02 |