Survey No. : RH2020-01 | Sample sent on : 2020 / 02 / 24 | Reporting deadline : 2020 / 03 / 02 |
This
Survey S3
has been used in 2013-2018 (n = 6; Survey / Sample No.) :
RH2009-07/S1 ; RH2012-03/S1 ; RH2013-09/S1 ; RH2014-01/S2 ; RH2016-10/S1 ; RH2017-01/S3 ; RH2020-01/S3 |
Intermediate Measurement Precision of G6PD Quantitative Test
Lab |
Survey : RH2020-01 / Sample 3 |
Period : 2009 ~ 20202 |
||||||
U/gHb |
SDI1 |
Mean |
n |
SD |
CV |
CVR |
||
RH01 |
11.5 |
0.9 |
10.0 |
7 |
1.73 |
17.3% |
2.79 |
|
RH02 |
11.5 |
1.1 |
10.9 |
7 |
0.54 |
5.0% |
0.80 |
|
RH04 |
9.7 |
-1.0 |
10.4 |
7 |
0.67 |
6.4% |
1.04 |
|
RH06 |
11.1 |
0.9 |
10.6 |
7 |
0.55 |
5.2% |
0.84 |
|
RH07 |
10.9 |
0.2 |
10.7 |
7 |
1.29 |
12.1% |
1.94 |
|
RH08 |
11.0 |
1.1 |
10.2 |
7 |
0.73 |
7.2% |
1.15 |
|
RH09 |
10.4 |
0.3 |
10.3 |
7 |
0.37 |
3.6% |
0.58 |
|
RH10 |
11.1 |
0.2 |
10.9 |
7 |
0.82 |
7.5% |
1.21 |
|
RH12 |
11.2 |
0.7 |
10.5 |
7 |
0.97 |
9.2% |
1.49 |
|
RH13 |
10.2 |
-1.0 |
10.6 |
7 |
0.42 |
4.0% |
0.64 |
|
RH14 |
11.2 |
1.6 |
10.4 |
7 |
0.50 |
4.8% |
0.78 |
|
RH19 |
10.0 |
-0.3 |
10.1 |
7 |
0.29 |
2.9% |
0.46 |
|
A0203 |
10.7 |
0.3 |
10.5 |
7 |
0.62 |
5.9% |
0.95 |
|
G026 |
10.8 |
0.1 |
10.7 |
7 |
0.76 |
7.1% |
1.15 |
|
CL001 |
10.3 |
0.4 |
10.1 |
7 |
0.47 |
4.7% |
0.75 |
|
CL002 |
9.5 |
-0.4 |
9.9 |
7 |
0.91 |
9.2% |
1.48 |
|
CL004 |
11.5 |
¢w |
¢w |
3 |
¢w |
¢w |
¢w |
|
CL014 |
11.0 |
0.6 |
10.6 |
6 |
0.64 |
6.0% |
0.97 |
|
CL015B |
10.5 |
-0.5 |
10.6 |
5 |
0.20 |
1.9% |
0.30 |
|
CL017 |
9.7 |
¢w |
¢w |
1 |
¢w |
¢w |
¢w |
|
Total participating laboratories = 20 |
||||||||
Median |
10.9 |
0.3 |
10.5 |
¢w |
0.63 |
6.0% |
0.96 |
|
Range |
9.5 ~ 11.5 |
1.0 ~ 1.6 |
9.9 ~ 10.9 |
¢w |
0.20 ~ 1.73 |
1.9% ~ 17.3% |
0.30 ~ 2.79 |
|
Mean of All participating laboratories CV
|
6.2% |
¢w |
3).
Survey / Sample No.¡GRH2009-07/S1 ; RH2012-03/S1 ; RH2013-09/S1 ; RH2014-01/S2 ; RH2016-10/S1 ; RH2017-01/S3 ; RH2020-01/S3 |
*
Intermediate Measurement Precision: Intermediate Measurement Precision¡FISO 3534-2:2006, JCGM 200:2012 ( VIM, 3rd ed. ) 2.23 |
** Mean and SD are calculated by Robust method according to ISO 13528, Algorithm A. |
*** n < 5¡Anot calculated Mean and SD |
Contact us
Address¡G
P.O. Box 26-624, Taipei, Taiwan, 10699¡@¡@Tel¡G+886-2-2703-6080
¡@¡@e-mail¡G
¡@Fax¡G
+886-2-2703-6070
Preventive
Medicine Foundation, Taipei, Taiwan
updated : 10 Jul, 2020