External Quality Assurance (EQA) Program

for

G6PD Quantitative Test in Taiwan


Survey No. : RH2017-01 Sample sent on : 2017 / 02 / 13 Reporting deadline : 2017 / 02 / 20
This Survey S2 has been used in 2013-2016 (n = 6; Survey / Sample No.) :
RH2013-11/S1 ; RH2014-07/S3 ; RH2015-10/S1 ; RH2016-07/S1 ; RH2016-10/S2 ; RH2017-01/S2

Intermediate Measurement Precision of G6PD Quantitative Test

Lab

Survey : RH2017-01 / Sample 2

Period : 2013 ~ 20172

U/gHb

SDI1

Mean

n

SD

CV

RH01

4.1
1.0
3.6
5
0.50
13.9%

RH02

3.8
-0.6
4.1
5
0.47
11.5%

RH04

4.0
0.0
3.7
5
0.19
5.1%

RH05

3.7
¢w
¢w
2
¢w
¢w

RH06

4.1
0.7
3.8
5
0.46
12.1%

RH07

3.9
0.0
3.9
5
0.32
8.2%

RH08

4.2
1.5
3.6
5
0.39
10.8%

RH09

3.9
1.2
3.7
5
0.17
4.6%

RH10

4.0
1.1
3.7
5
0.27
7.3%

RH12

4.1
1.3
3.4
5
0.53
15.6%

RH13

4.2
0.7
3.9
5
0.44
11.3%

RH14

4.4
1.3
3.7
5
0.55
14.9%

RH18

3.5
0.0
3.5
5
0.30
8.6%

RH19

3.3
-0.7
3.5
4
0.30
8.6%

A0189

3.4
-0.9
4.1
5
0.79
19.3%

A0203

3.7
0.2
3.6
5
0.52
14.4%

G026

3.4
-0.8
3.7
5
0.39
10.5%

CL001

4.0
0.2
3.9
5
0.58
14.9%

CL002

4.1
1.0
3.8
5
0.31
8.2%

CL014

3.2
-1.0
3.5
5
0.31
8.9%

CL015B

3.7
0.7
3.4
5
0.44
12.9%

Total participating laboratories = 21

Median

3.9
0.4
3.7
¢w
0.4
11.1%

Range

3.2 ~ 4.4
-0.97 ~ 1.54
3.4 ~ 4.1
¢w
0.17 ~ 0.79
4.6% ~ 19.3%

 

1). Standard Deviation Index (SDI) = (Individual Result - Individual Mean) / Individual SD
*Intermediate Measurement Precision¡FISO 3534-2:2006, JCGM 200:2012 ( VIM, 3rd ed. ) 2.23
** Mean and SD are calculated by Robust method according to ISO 13528
*** n < 5, not calculated Mean and SD.

 


Contact us
Address¡G P.O. Box 26-624, Taipei, Taiwan, 10699¡@¡@Tel¡G+886-2-2703-6080
  ¡@¡@e-mail¡G
¡@Fax¡G +886-2-2703-6070

     
                  Preventive Medicine Foundation, Taipei, Taiwan