External Quality Assurance (EQA) Program

for

G6PD Quantitative Test in Taiwan


Survey No. : RH2014-10 Sample sent on : 2014/09/22 Reporting deadline : 2014/09/29
This Survey S1 has been used in 2008-2014 (n = 6; Survey / Sample No.) :
RH 2010-07/S3 ; RH 2011-01/S1 ; RH 2011-09/S3 ; RH 2012-01/S1 ; RH 2013-09/S2 ; RH 2014-10/S1 ;


Intermediate Measurement Precision of G6PD Quantitative Test

Lab

Survey : RH2014-10 / Sample 1

Period : 2010 ~ 20142

U/gHb

SDI1

Mean

n

SD

CV

RH01

10.8
-1.0
12.9
6
2.1
16.3%

RH02

12.7
-0.1
12.8
6
0.7
5.5%

RH03

8.2
-7.0
14.5
53
0.9
6.2%

RH04

13.0
-0.7
13.6
6
0.9
6.6%

RH06

13.0
0.0
13.2
6
0.5
3.8%

RH07

12.7
-1.0
12.9
6
0.2
1.6%

RH08

12.7
-1.0
12.9
6
0.2
1.6%

RH09

11.9
-1.0
13.4
6
1.5
11.2%

RH10

12.9
-0.4
13.1
6
0.5
3.8%

RH12

12.0
-0.8
12.4
6
0.5
4.0%

RH13

14.4
0.4
13.8
6
1.5
10.9%

RH14

13.0
-0.6
13.3
6
0.5
3.8%

RH18

14.0
0.0
13.5
6
0.4
3.0%

RH19

13.6
0.6
13.1
6
0.9
6.9%

A0189

12.5
-1.2
13.2
6
0.6
4.5%

A0203

11.2
-1.9
13.1
6
1.0
7.6%

G026

13.7
0.6
13.0
6
1.1
8.5%

CL001

13.6
0.8
12.9
6
0.9
7.0%

CL002

13.5
0.8
12.9
6
0.8
6.2%

B001

12.5
-0.3
12.9
6
1.3
10.1%

CL014

13.4
0.0
13.4
6
0.7
5.2%

CL015B

13.1
13.1
2

Total participating laboratories = 22

Median

13.0
-0.4
13.1
0.8
6.2%

Range

8.2 ~ 14.4
-7.0 ~ 0.8
12.4 ~ 14.5
0.2 ~ 2.1
1.6% ~ 16.3%
1). Standard Deviation Index (SDI) = (Individual Result - Individual Mean) / Individual SD
3). Because the result was assay by other lab, so the IMP result for RH03 is not included.
*Intermediate Measurement Precision¡FISO 3534-2:2006
** Mean and SD are calculated by Robust method according to ISO 13528:2005

 


Contact us
Address¡G P.O. Box 26-624, Taipei, Taiwan, 10699¡@¡@Tel¡G+886-2-2703-6080
  ¡@¡@e-mail¡G
¡@Fax¡G +886-2-2703-6070

     
                  Preventive Medicine Foundation, Taipei, Taiwan

updated : 17 Jul, 2015


[ Home ]